Problem:
GE wall oven unexpectedly shows “150° Probe” on the lower oven display, even when no probe is inserted or no cooking cycle is active.
Possible Causes:
-
Faulty or shorted meat probe jack
-
Moisture or debris inside probe connector
-
Damaged or misaligned probe receptacle wiring
-
Software or control board glitch
-
Faulty main control board interpreting ghost probe signal
Tools Required:
-
Screwdriver
-
Multimeter
-
Needle-nose pliers
-
Rubbing alcohol + cotton swab
-
Flashlight
Step-by-Step Procedure:
1. Power Cycle the Oven
-
Turn off power at the breaker for 5 minutes.
-
Restore power and monitor if the “150° Probe” display reappears.
-
If problem is temporarily resolved, likely a control issue.
2. Inspect Meat Probe Jack
-
Open the lower oven.
-
Locate the probe receptacle inside the oven cavity (usually rear or side wall).
-
Check for:
-
Debris
-
Moisture
-
Grease or food particles
-
-
Clean with cotton swab and rubbing alcohol.
-
Allow to dry completely.
3. Check for Phantom Probe Detection
-
Some ovens detect a probe when the jack is physically damaged or shorted internally.
-
Use a multimeter to test the two probe jack terminals for continuity:
-
With no probe inserted, there should be no continuity.
-
If there is continuity, the jack is internally shorted and must be replaced.
-
4. Disconnect Probe Wiring (Test)
-
Turn off power to the oven.
-
Remove oven from wall if necessary (2–4 screws usually at front frame).
-
Access rear or top control compartment.
-
Locate the probe wire harness going from the cavity to the control board.
-
Disconnect the probe wiring harness.
-
Restore power and check if “150° Probe” message disappears.
-
If yes, replace the probe jack assembly.
-
If not, suspect control board issue.
-
5. Check Control Board
-
If the display still shows “150° Probe” with the jack disconnected, the control board may falsely interpret probe input.
-
Inspect control board for:
-
Burn marks
-
Loose connections
-
Corroded pins
-
-
Replace main control board if fault persists.
Notes:
-
This fault can also intermittently occur after self-clean cycles due to thermal expansion affecting probe jack or wiring.
-
Moisture in the cavity (from steaming/roasting) may also cause false detection until it evaporates.
Replace If Necessary:
-
Probe jack assembly – if shorted or corroded
-
Control board – if false signal persists with probe disconnected